D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-3
3.) Using both hands, pull the Bearing Handle slightly apart to release it from the system.
4.) Remove the Bearing Handle.
8-2-1-5 Bearing Handle Installation Procedure
1) Place a new Bearing Handle in position, using both hands to pull the Bearing Handle slightly apart
in order to slot the ends into the handle holes on the Vivid q N casing.
2.) Carefully fasten each captive screw on either side of the Bearing Handle to secure the handle in
position, by turning 1/4 turn in a clockwise direction.
NOTE: When tightening the captive screws, do not use excessive force do not apply more than
1/4 turn; overtightening will damage the screw thread.
3.) Proceed to perform the "Bearing Handle Test" on page 4-19.
Figure 8-3 Bearing Handle Removed

Table of Contents