IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-123
11.)Press <F10> to Exit Saving Changes as shown in Figure 8-128.
12.)Reboot the system and check that the Supervisor PASSWORD was installed properly.
13.)Proceed to perform the following functionality tests:
- "Software Configuration Checks" on page 4-39
- "Calibration Tests" on page 7-25
- "Full System Test" on page 7-10
Figure 8-128 BIOS Exit Screen