GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
3-134 Section 3-4 - Preparing for Installation
PRELIMINARY
1.) Select QueryRetrieve from the Selected devices list.
2.) Select Check.
- If the test passes, a pop-up dialog is displayed, see left illustration in Figure 3-157.
Select OK to continue. A green check mark to the left of the Selected devices indicates that
the test passed successfully.
NOTE: To use Query/Retrieve, unhide the Q/R Dataflow, then select Query Retrieve Dataflow in the Archive
tab. Use Name, Date, etc. to filter results then select Query button. Select patient, then select Retrieve
button.
- If the test fails, a pop-up warning is displayed, see left illustration in Figure 3-158.
Figure 3-156 Check the DICOM Query/Retrieve Workflow
Figure 3-157 Test (Check) Passed