EasyManua.ls Logo

GE Vivid S5 N - 8-4-8 Extended Keyboard Replacement Procedure (New-type Operator Panel)

GE Vivid S5 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-79
PRELIMINARY
8-4-8 Extended Keyboard Replacement Procedure (New-type Operator Panel)
NOTE: The procedures described in this section are applicable to all Vivid S5 N / Vivid S6 N systems supplied
with the new-type Operator Panel.
NOTE: For Vivid S5 N / Vivid S6 N systems with the older-type Operator Panel, refer to the instructions
provided in “Extended Keyboard Replacement Procedure (Old-type Operator Panel)” on page 8-82.
NOTE: The extended keyboard (P/N 5430996), shown in Figure 8-94, is only compatible with the new
IR keyboard (P/N 5376245)
NOTE: A spare On/Off switch and Screen sensor is supplied with the kit.
Figure 8-93 Vivid S5 N / Vivid S6 N - New-type Operator Panel
CAUTION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to “Battery Removal Procedure” on page 8-116.

Table of Contents