NOTE: This procedure is applicable to MODBR964 version.
NOTE: To reset the BIOS password, one can remove the real time battery for a period longer than 10 secs,
then reconnect it. (It is recommended to replace the battery once every 2 years).
1) While booting, press F2 “Config / Diag” on alphanumeric keyboard
2) Type the password “marlin04”.
3) Exit and save presets.
4) Select <Exit> section and load “Setup Defaults”.
5) Select <Main> section.
6) Set correct system time and Date. Set Date using next format MM/DD/YYYY
7) Check that IDE primary master is detected.
8) Select <Advanced> section
9) Perform the following changes in the “Advanced”:
10) Perform the following changes in the <Security>:
11) Perform the following changes in the <Power>
Option Value
\Advanced\PnP OS Installed No
\Advanced\Secured Setup Configuration No
\Advanced\PCI configuration\Assign IRQ to PCI VGA No
\Advanced\IO Device Configuration\Serial Port A Enabled
\Advanced\IO Device Configuration\Serial Port B Enabled
\Advanced\IO Device Configuration\OnChip Audio Device Enable
Option Value
\Security\Set Supervisor Password marlin04
\PNP/PCI\IDE DMA transfer access Enabled
Option Value
\Power\Wake Up Events\IRQ1 Disabled
\Power\Wake Up Events\IRQ12 Disabled
\Power\Wake Up on LAN Disabled