ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 3 - System Setup 3-97
PRELIMINARY
Automatically send images to DICOM PACS after End Exam? Y/N
DICOM SR? Y/N
Note: If the answer to some of the above is Yes, perform the required steps in the procedure below
to select the appropriate dataflow according to your needs. If the answer to all of the above is
Yes, perform all the steps below:
1.) Access the Configuration > Connectivity settings and click on the Dataflow tab.
2.) Select the dataflow named Worklist/Local Archive - DICOM Server/Int HD - see Figure 3-110.
3.) Under Selected devices > Inputs, select Worklist:
Figure 3-110 Selecting Required Dataflow
Figure 3-111 Selecting Worklist