GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 3 - System Setup 3-141
PRELIMINARY
3-7-8 eVue Dataflow Configuration
NOTE: MPEGVue Player is intended for secondary review and patient consultation.
3-7-8-1 Introduction
It is possible to configure the Vivid S5 N or Vivid S6 N ultrasound scanner to allow the user to send
images or loops to a remote PC, in almost real time. When pressing Store, the loop will save not only
to the local archive (in raw-data format), but will also be transmitted to a designated PC (“e-Vue PC”)
on the network (in compact MPEG format), and will appear at the review PC within a very short time.
This feature proves very useful in the OR scenario, where consultation with another doctor in the
hospital in real-time is required.
3-7-8-2 Preparations
Before configuring the Vivid S5 N or Vivid S6 N for eVue dataflow, it is necessary to decide upon a
specific PC in the hospital as being the e-Vue PC station. The service technician then configures the
Vivid S5 N or Vivid S6 N to send images to that specific e-Vue PC, as described below.
3-7-8-3 Configuring the Vivid S5 N or Vivid S6 N for eVue Dataflow
1.) Verify which PC in the hospital/institute has been designated as the e-Vue station.
2.) Make sure the Vivid S5 N or Vivid S6 N is connected to the Network via cable or wireless.
3.) On the Vivid S5 N or Vivid S6 N scanner, press the Config button.
The System Configuration window opens.
4.) Click Connectivity.
The Connectivity dialog box opens.
5.) Select the Dataflow Tab, as shown in Figure 3-163 on page 3-142.