ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-69
PRELIMINARY
7-4-6 Probe Select Board Diagnostic Tests (cont’d)
8) Select the second report (Signal RMS); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-40 "View Test
Log - ACW Matrix Test - Signal RMS" on page 7-69.
.
9.) Use the tools provided to scroll through the channels and view the results, as previously described.
10) When done, trackball to the Close button and press Set to close the View Report window.
11) Select the third report (Signal SNR); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-41 - View Test
Log - ACW Matrix Test - Signal SNR.
Figure 7-40 View Test Log - ACW Matrix Test - Signal RMS