ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-95
PRELIMINARY
7-5-1 Accessing the Monitoring Diagnostic Test Options (cont’d)
15.)Check that the sub-test passed and that result values are within the pre-defined Upper and Lower
limits, as shown in Figure 7-63 "View Test Log - Full System Monitoring Test - Battery Parameters"
on page 7-95.
16.)Select the fifth report (Tx Voltages); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in Figure 7-64 "View Test
Log - Full System Monitoring Test - Tx Voltages" on page 7-96.
Figure 7-63 View Test Log - Full System Monitoring Test - Battery Parameters