ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-111
PRELIMINARY
8-6-1 Opening and Removing the Cabinet Cage Assembly (cont’d)
7.) Unfasten the two knurled (ridged) screws on the Docking Connector and release the Docking
Connector from the Back End (Figure 8-134
).
8.) Cut the cable tie securing the KB Data cable to the battery bracket.
9.) Using a flat screwdriver, release the two connectors that connect the KB Data cable to the KB outlet
on the Back End and to the ECG USB connector on the one side; on the other side to the bundle
of cables connected to the DIB.
Figure 8-134 Releasing the Docking Connector
Figure 8-135 Location of Cable Tie
Figure 8-136 Releasing Connectors on the KB Data Cable
Docking Connector
Cable Tie