ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-143
PRELIMINARY
8-6-6 ECG Module Assembly Replacement Procedure (cont’d)
4) Turn the ECG Module assembly (with bracket attached) over, ready to place it in the correct location
as shown in Figure 8-176.
5.) Carefully slide the assembly and fastening bracket sideways to the right until it is seated in the
correct position. Make sure the securing bracket holes are properly aligned - Figure 8-176.
6) Fasten the three screws on the bracket to secure the ECG Module assembly to the cabinet cage
assembly (refer to Figure 8-171 on page 8-141).
7) Return the cabinet cage assembly to the system, reconnect the cables (previously disconnected),
close the cabinet cage door and fasten securely with the two latches.
Refer to the 8-6-1-4 "Cabinet Cage Assembly Closing and Installation Procedure" on page 8-114.
8) Return the front cover, as described in the 8-2-6-4 "Front Cover Installation Procedure" on page 8-
19.
9) Return the left and right side covers, as described in the 8-2-3-4 "Left Side Cover Installation
Procedure" on page 8-7 and 8-2-2-4 "Right Side Cover Installation Procedure" on page 8-5.
10.)Turn ON power to the system.
Figure 8-175 Return the ECG Module Assembly and Bracket
Figure 8-176 Sliding the ECG Module Assembly and Bracket into Position