ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-162 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
4.) Attach the TR32 Boards to the RFI Board by gently pressing the two parts together as shown in
Figure 8-203.
5) Return the TR32 and RFI Board assembly - together with the Heat Sink pad - to the appropriate
position in the cage assembly - Figure 8-204.
Note: When performing this step, carefully slide the Heat Sink into position.
6) Reconnect the BEP-to-RFI flex cable to the cable connector on RFI Board, as shown in
Figure 8-193 on page 8-157.
7) Reconnect the power cable to the RFI Board - see Figure 8-193.
Figure 8-203 Attaching the TR32 Boards to the RFI Board
Figure 8-204 Refitting TR32 and RFI Assembly (incl. Thermal Pad) into the Cage Assembly
Thermal Pad
TR32 and RFI