ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 10 - Care and Maintenance 10-5
PRELIMINARY
Table 10-2 Customer Care & Maintenance Task Schedule
Service at Indicated Time Daily Weekly Monthly Annually Notes
Clean Probes •
Clean Gel Cup •
Clean Probe Holders •
Clean Air Filter •
More frequently if necessary, depending on
the environment
Inspect AC Mains Cable
*• *Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console •
Clean Monitor •
Clean Keyboard (KB) •
Clean Trackball •
Inspect Wheels, Casters, Brakes and
Swivel Locks
*•* Mobile Unit Check Daily
Check Control Panel Movement
*•* Mobile Unit Check Daily
Console Current Leakage Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Peripheral
Current Leakage Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Surface Probe
Current Leakage Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Endocavity Probe Current Leakage
Checks
As prescribed in the Probe Manual.
Transesphongeal (TEE) Probe
Current Leakage Checks
As prescribed in Probe Manual.
Surgical Probe
Current Leakage Checks
As prescribed in Probe Manual.
Measurement Accuracy Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
Probe/Phantom Checks
•
Only recommended if required by your
facility’s QA program.
Functional Checks
•
Also after corrective maintenance,
or as required by your facility’s QA program.
BIOS Battery
*
It is recommended to replace the BIOS
Battery once every two years.
System Battery
*
It is recommended to replace the system
Battery once every two years.