EasyManua.ls Logo

GE Vivid S70 - Page 38

GE Vivid S70
709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Safety
2-6 Vivid S70 / S60User Manual
BC092760-1EN
01
Neonatal trans-cranial & spinal scanning
TI time Note
0.0–0.7 Unlimited Monitor TIC. MI>0.7 should be used with caution in the presence of
contrast agents.
0.7–1.0 < 60 min
1.0–1.5 < 30 min
1.5–2.0 < 15 min
2.0–2.5 < 4 min
2.5–3.0 < 1 min
Neonatal general and cardiac scanning
TI time Note
0.0–1.0 Unlimited Monitor TIB. MI>0.7 should be used with caution in the presence of
contrast agents.
1.0–1.5 < 120 min
1.5–2.0 < 60 min
2.0–2.5 < 15 min
2.5–3.0 < 4 min
Adult trans-cranial scanning
TI time Note
0.0–1.0 Unlimited Monitor TIC. MI>0.7 should be used with caution in the presence of
contrast agents.
1.0–1.5 < 30 min
1.5–2.0 < 15 min
2.0–2.5 < 4 min
2.5–3.0 < 1 min
General Abdominal, Peripheral Vascular and other scanning (excluding the eye)
TI time Note
0.0–1.0 Unlimited Monitor TIB or TIC if bone closer than 1 cm, TIS if no bone is in the
image. MI>0.7 should be used with caution in the presence of
contrast agents.
1.0–1.5 < 120 min
1.5–2.0 < 60 min
2.0–2.5 < 15 min
2.5–3.0 < 4 min

Table of Contents

Other manuals for GE Vivid S70

Related product manuals