IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-69
PRELIMINARY
9.) Remove the cable holder from the upper arm - see Figure 8-86 below.
NOTE: Before threading the cables through the aperture in the OIPIO platform, the "up-down lever shaft" must
be removed from obstructing the cable aperture as shown in - Figure 8-87 below.
10.)Remove the "up-down lever shaft" securing screw and move the shaft aside.
Figure 8-86 Cable Holder on Upper Section of Lower Arm Cover
Figure 8-87 Releasing Up-Down Lever Shaft