Connectivity setup
Vivid T9/Vivid T8 – Basic Service Manual 3-49
5795591-100 English Rev. 3
3-9-4-1 DICOM Server IP address on the ultrasound
system(continued)
Figure 3-33. Select DicomWorklist > Properties
7. In the Properties dialog, select the IP-address dialog
pull-down menu to select the Worklist Server.
It is not possible to change the setting in the IP-address filed
by editing it.
To change the IP-Address settings, refer to 3-9-2 ‘Set the
Remote Archive’s Network Information’ on page 3-44.
• Enter the DICOM server’s AE Title. This entry is case
sensitive and must match exactly.
• Enter the DICOM server’s Port No.(Port number).
• For some DICOM servers, the default Time-out setting
(30) is too low.
• Select OK to close the Worklist properties dialog and
save changes.
Figure 3-34. The DicomWorklist Properties Window