Diagnostics/Troubleshooting
7-24 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
7-4-5-6 FRU Status
FRU Status displays a summary of the FRUs being monitored
and the details for each FRU.
To access this page, under System Status, select Temperature
and then select FRU Status.
Figure 7-11. FRU Status
This table shows all the elements available on FRU Status with
descriptions.
Table 7-5: FRU Status
Element DESCRIPTION
FRU Status Summary Number of the FRUs being monitored.
FRU Status Details
Name Name of the FRU.
• ETX3 - Represents the ETX 64 transmitter board.
• ETX4 - Represents the ETX 128 transmitter board.
Status Status of the FRU. Valid values are:
• Normal - Indicates that the FRU is operating within the
allowable range.
• Warning - Indicates that the FRU is operating close to
the limit of the allowable range.
• Error - Indicates that the FRU is operating outside the
allowable range.
Low Level Warning Number of low temperature warnings.
High Level Warning Number of high temperature warnings.
Low Level Errors Number of low temperature errors.
High Level Errors Number of high temperature errors.