Common Diagnostics
Vivid T9/Vivid T8 – Basic Service Manual 7-37
5795591-100 English Rev. 3
7-5-1 Remote Access(continued)
4. Enter Otpions to add service option key.
NOTE: Please delete the added service option key when your
service work is done.
Figure 7-22. Add Service Option Key
5. Enter Utility-> Disruptive Mode Utility, select Enable.
Figure 7-23. Enable Disruptive Mode
6. If we want to see the Scan screen or have control of the unit,
we have to go to Virtual Console Observation and start it.
Enter Utility-> Virtual Console Observation, select Start.
Figure 7-24. Start VCO