8-4-9 Option Strings Check
NOTE: After the system software loading completion, please check the
option strings to ensure that the options are activated and
working.
1. Reboot the system.
2. Press Config (F2) on the control panel.
3. Select Admin -> System Admin.
4. Ensure that all the installed option keys are displayed and
the status of Options are valid.
The status “Valid” means the option keys are activated
and working.
The status “disabled” means the option keys are not
activated and not working. Check if the option is
installed and if the serial number and option key are
correct.