GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 4 - Functional Checks 4-3
4-3-1-1 Scanner Power On (cont’d)
5.) Press the ON/OFF Standby switch left below the control panel.
When the ON/OFF Stand-By switch left below the control panel is pressed, the System (including the
Back-end Processor) starts and the software code is distributed to initiate the scanner.
Depending on the BIOS-Version no status messages are displayed during this process.
Boot up time is about 2 minutes.
NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with
the Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
4-3-2 Power Off / Shutdown
NOTE: After turning off a system, wait at least ten seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
4-3-2-1 Scanner Shutdown
1.) Press the ON/OFF Standby switch left below the Control Panel.
2.) Switch OFF the Circuit Breaker at the rear of the system.
3.) Disconnection of the Mains Power Cable is necessary. For example: Relocating the scanner.
Unscrew the 2 screws and remove the pull-out protection to disconnect the cable from the system,
or unplug the cable from the wall socket.
Figure 4-2 ON/OFF Standby Switch
Figure 4-3 Circuit Breaker, Protection and Power Cable on back of Voluson® 730Pro / 730ProV
ON/OFF Standby switch
Circuit Breaker
Mains Power Cable
pull-out Protection
2 screws