GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
4-30 Section 4-4 - Functional Checks
4-4-11-1 Formatting Media (cont’d)
By selecting DVD/CD the unit displays the “Erase DVD/CD” window as shown in Figure 4-19.
1.) Select the “Erase Mode” and click the [OK] button to start the process.
2.) During erasing the Ultrasound system displays following message.
3.) Press the EXIT
key on the control panel to return to the Scan Mode.
4-4-11-1-1 Adjusting the Write Speed of the DVD/CD Recorder
1.) Click the SETTINGS
icon on the left side of the Sonoview screen.
2.) Select
DVD/CD RECORDER from the tool bar.
3.) Choose the “Write Speed” from the corresponding pop-up menu.
NOTICE
It is highly recommended to use the complete erase mode, to avoid problems with the CD+(R)W!
When using a DVD+(R)W, the complete erase mode is not possible.
Figure 4-19 Erase DVD/CD Window
Figure 4-20 Erasing disk
Figure 4-21 Settings of the DVD/CD Recorder
Settings
icon