GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
4-32 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-1 Save User Settings Only (Application Settings)
The User Settings contains:
•User Programs
•Auto Text
3D/4D Programs
1.) Insert a DVD/CD+(R)W or MO (Magneto-Optical Disk) into the drive.
2.) Press the UTILITIES
key on the control panel. The menu area changes to the Utilities menu.
3.) Select the SYSTEM
item from the menu area to activate the setup desktop screen.
4.) Select the
BACKUP page.
5.) Click the SAVE
button of the “User Settings Only” group.
6.) Choose the media (e.g., DVD/CD+RW) and click the SAVE
button.
7.) Select the NEW FILE...
key and enter a file name (without extension).
8.) Click the OK
key to start the process. When the saving has been completed, click OK.
Figure 4-23 User Settings Only - Save window
Choose the media
and
then click Save