EasyManua.ls Logo

GE Voluson 730Pro - CW Mode (Continuous Wave Doppler)

GE Voluson 730Pro
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
4-34 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-2 Load User Settings Only (Application Settings) (cont’d)
7.) Select the Complete Backup (marked blue; see: Figure 4-25) and click the >> button to copy the
Complete Backup into the Load Data field.
8.) Click the LOAD
button to start the loading procedure of the complete backup into the system.
NOTE: Also only parts of a User Settings Only “Backup” can be loaded into the database to overwrite,
restore, copy, etc.... the database in the system.
1.) Click the +
sign to open the content tree.
2.) Click the >>
button to copy the selected item into the Load Data field.
3.) To return selected items from the Load Data field to Backup Data field select the <<
button.
4.) Click the LOAD
button to start loading procedure of the selected Backup item into the system.
Figure 4-26 Load only parts of the Backup
sign to open the
content tree
>>
button
<< button
Load button

Table of Contents

Other manuals for GE Voluson 730Pro