GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
4-34 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-2 Load User Settings Only (Application Settings) (cont’d)
7.) Select the Complete Backup (marked blue; see: Figure 4-25) and click the >> button to copy the
Complete Backup into the Load Data field.
8.) Click the LOAD
button to start the loading procedure of the complete backup into the system.
NOTE: Also only parts of a User Settings Only “Backup” can be loaded into the database to overwrite,
restore, copy, etc.... the database in the system.
1.) Click the +
sign to open the content tree.
2.) Click the >>
button to copy the selected item into the Load Data field.
3.) To return selected items from the Load Data field to Backup Data field select the <<
button.
4.) Click the LOAD
button to start loading procedure of the selected Backup item into the system.
Figure 4-26 Load only parts of the Backup
sign to open the
content tree
>>
button
<< button
Load button