GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 4 - Functional Checks 4-41
Section 4-6
Software Configuration Checks
Press the UTILITIES key and select the SYSTEM item from the “Utilities” menu - which is shown in the
menu area on the left side of the screen. The System Setup desktop offers different pages to check:
Press the UTILITIES
key and select the MEASURE item from the “Utilities” menu - which is shown in
the menu area on the left side of the screen.
The Measure Setup desktop offers three different pages to check.
Table 4-10 Software Configuration (System Setup) Checks
Step Task Expected Result(s)
1
General: Check Date and Time setting Date and Time are correct
2
General: Check that Location (Clinic Name) is correct Location Name is correct
3
General: Check Language settings desired Language is displayed
4
User Setting: Check all the User Settings settings assigned as desired by the customer
5
User Settings: Check settings of the “Patient ID” field settings assigned as desired by the customer
6
Peripherals: Check assignment of Printer keys
Print A and Print B keys are assigned as desired by the
customer
7
Peripherals: Check assignment of Foot switch
Foot Switch left and right are assigned as desired by the
customer
8
Peripherals: Check Save Destination assignment assigned as desired by the customer
9
Peripherals: Check 2D Save key assignment assigned as desired by the customer
10
Peripherals: Check Remote Print A, Remote Print B,
Report Printer and DICOM Print Job assignment
settings assigned as desired by the customer
11
Option: Check that all options are set up correct D = Demo , I = Inactive , P = Permanent
12
Service: Check the Video Norm standard as described in
Section 3-8-3-4 on page 3-42
settings assigned as required for the country
13
Network:
Check DICOM, Sonoview and Network configuration
settings assigned as desired by the customer
NOTICE
Parameters and possible adjustments depend on the selected Application.
To view, add, delete, reorder, edit or when creating a new parameter (in the
MEASURE & CALC page), it
is very important that all items are chosen correctly and that the relevant item is highlighted.
For further information refer to Chapter 18 in the Basic User Manual of Voluson® 730Pro / 730ProV.
Table 4-11 Measurement Setup Checks
Step Task Expected Result(s)
11
Measure & Calc:
Check all settings for all applications
setting assigned as desired by the customer
2
Application Parameters:
Check all settings for all applications
setting assigned as desired by the customer
3
Global Parameters: Check all settings setting assigned as desired by the customer