EasyManua.ls Logo

GE Voluson 730Pro - CFM Mode (Color Flow Mode)

GE Voluson 730Pro
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 4 - Functional Checks 4-43
Section 4-8
Mechanical Function Checks
4-8-1 Rotation of the Control Console
4-8-2 Brakes and Direction Locks
Figure 4-37 locking lever under Control Console
Table 4-14 Rotation of the Control Console
Step Task Expected Result(s)
1
Pull the locking lever under the Control Console forward,
grasp it at the front grip and rotate the console.
It is possible to rotate the Control Console up to 30° to
the right.
WARNINGWARNING
Do not put your hand between the control console and the main unit when moving the
console to the 0 position: Danger of injuries!
Figure 4-38 Front Wheel with Brake
Table 4-15 Brakes and Direction Lock
Step Task Expected Result(s)
1
Flap the foot rest up and press the release brake pedals on
the front wheels.
The front wheels are engaged / disengaged for
transportation.
locking lever
for locking and unlocking
the control console
Brake
pedal
steer able
Front Wheel

Table of Contents

Other manuals for GE Voluson 730Pro

Related product manuals