GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
5-54 Section 5-15 - Service Page
5-15-3-3 Service Viewer
Provides common information about System Temperature, Probes, Working hours of system
components and probes.
1.) Select the SERVICE VIEWER
button to get access to the E-Service page.
5-15-3-4 Delete all Patients
1.) Click the DELETE ALL PATIENTS...
button. Following WARNING message appears on the screen.
5-15-3-5 Export System Data
Select the EXPORT SYSTEM DATA
button on the “Service Tools” page to export dump-files and text
files, the full Service Database informations about probes, boards, Software, Options, Service Actions
and the Event Log File to the DVD/CD Drive (or the optional MO Drive).
Operation see: Section 7-4-2 "Export Log’s and System Data" on page 7-5.
5-15-3-6 Maintenance Report
Any modification upgrade and maintenance action should be entered in this report to get a history of all
service actions.
Figure 5-39 Kretztechnik E-Service
Figure 5-40 Warning message
WARNINGWARNING
If you select the YES button, all patients data, studies, images and measure report data
will be deleted permanently from the hard disk and cannot be recovered!