GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
5-56 Section 5-15 - Service Page
The Monitor Test main menu appears on the screen.
3.) Open the desired test page: select the respective icon and press the
RIGHT/LEFT TRACKBALL key.
4.) To return to the Main menu, press the
UPPER TRACKBALL key.
5.) To exit the Monitor Test program, click the “Exit” icon.
5-15-3-8 Printer
Installation of the Line Printer is possible without entering the Windows Desktop.
Operation see: 3-7-3 "Printer Installation manually" on page 3-28.
5-15-3-9 Update
5-15-3-9-1 NLS
Not for use in the field!
5-15-3-9-2 EUM
is for updating the Electronic User Manual
Operation see: Section 8-6 "Electronic User Manual (EUM) Upgrade Procedure" on page 8-11.
5-15-3-9-3 UIS
is for updating the Ultrasound Application Software
Operation see: Section 8-2-5 "Software - Installation Procedure (via Service Page)" on page 8-5.
5-15-3-9-4 SSW
is for updating the Service Software
Operation see: Section 8-5 "Service Platform (SSW) Upgrade Procedure" on page 8-9.
Figure 5-43 Monitor Test - Main menu
WARNINGWARNING
Only accessories explicitly recognized and released by the system manufacturer GE
Medical Systems - Kretztechnik may be used in connection with the system.
EXIT