GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
7-6 Section 7-4 - Screen Captures and Logs
7-4-2-1 Export System Data (by pressing the ALT + D key)
ALT
+ D uses “Full Backup” to gather data from the system. In addition it creates one screenshot
(Alt-D.bmp) of the point in time when ALT
+ D was pressed.
The main use is when R&D or OLC need detailed information about the system (e.g., when
experiencing strange behaviour or when the problem should be investigated by R&D).
It is not intended to replace or enhance the existing Full Backup functionality.
Data can be stored on the hard disk (D:\export\fullbackup\fb1), or you can export them to DVD/CD
(including the D:\export folder, which contains logs and dump-files; for details see: Section 7-4-2-2-1).
NOTE: Successive ALT
+ D keystrokes overwrite existing snapshots at destination HDD.
1.) Press the ALT
+ TAB key on the keyboard. simultaneously.
2.) Select the destination of the “System state” backup among HDD or DVD/CD.
3.) Select the NEXT
button to start the backup process.
After saving the data, the Voluson® 730Pro / 730ProV reboots and the application starts again.
Figure 7-3 select destination for “System state” backup