GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-7
7-4-2-2 Export Log´s and System Data (via Service Page)
1.) Press the UTILITIES
key, select the SYSTEM item from the menu area.
2.) Select the
SERVICE page on the screen. The “password window” appears automatically.
3.) Enter the password SHE and click the ACCEPT
button to display the Service Tools window.0
4.) Click on the EXPORT SYSTEM DATA...
button, to export dump-files (see: Section 7-4-2-2-1), text
files, the Event Log File, the full Service Database and information about probes, boards, software,
options and service actions to DVD/CD Drive (or the optional MO drive).
7-4-2-2-1 Dump-file
Every time an error message like Figure 7-5 is produced, a dump-file and a text file containing the
error dump and the error message are created in D:\export. Up to 20 dump files are stored there.
After clicking on OK
the system reboots automatically.
Figure 7-4 Service Tools window
Figure 7-5 system has encountered a problem
Export System Data
important
dump-file
informations