GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
7-12 Section 7-6 - Minimum Configuration to Boot/Scan
7-6-2 Minimum Configuration to Scan
Following cables must be connected to scan; see: Figure 7-14
a.) PS2 (connector for Mouse/Keyboard)
* PS2 Adapter is ONLY used, when the currently installed PC-Board is an IBASE Standard Board Computer.
If the PC-Board is a Kontron SBC, the PS2 cable [a] has to be connected directly to the PC-part of the GEF-box.
b.) UI Disp. (connector for User Interface Display - NOT used at Voluson® 730Pro / 730ProV)
c.) VGA (Monitor)
d.) Console
e.) CPN (Primary Power Supply)
f.) Standby switch
g.) Monitor (Power Supply)
Connect mains power cable to the system and to an appropriate mains power outlet.
Connect a probe, boot up the system and start an User program.
Figure 7-14 cable- minimum configuration
a
b
c
d
f
g
e
*
PS2
Adapter
not used on
Voluson® 730Pro /
Voluson® 730Pro V