GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-21
7-7-8 DVD/CD+(R)W Troubleshooting (DVD/CD Drive)
1.) Insert an empty DVD/CD+(R)W into the Drive.
2.) Enter “Sonoview” by pressing the SONOVIEW
key on the control panel; see Figure 7-22.
3.) Click the “Open” icon to display the list of exams.
4.) Select exam(s) and backup them to DVD/CD+(R)W.
5.) Choose “DVD/CD” Drive.
6.) The images, which you have choosen during backup should be visible.
7-7-9 MOD Troubleshooting
1.) Insert an empty MO into the Drive.
2.) Enter “Sonoview” by pressing the SONOVIEW
key on the control panel; see Figure 7-22.
3.) Click the “Open” icon to display the list of exams.
4.) Select exam(s) and backup them to MO.
5.) Choose “MO” Drive.
6.) The images, which you have chosen during backup should be visible.
Figure 7-22 To backup exams to DVD/CD+(R)W resp. MO-Disk
NOTICE
In case of any problems, check status of the Diagnostic LED´s and voltages of the AUX Supply
connector (= Disk Drive Module power connector) on the backpanel of the GEF-Box;
see: Section 7-3 "Check Points Voltages" on page 7-4
NOTICE
In case of any problems, check status of the Diagnostic LED´s and voltages of the AUX Supply
connector (= Disk Drive Module power connector) on the backpanel of the GEF-Box;
see: Section 7-3 "Check Points Voltages" on page 7-4
1. Press Sonoview key 2. Click “Open” icon
and backup exams
3. After Backup,
choose DVD/CD
(resp. MO)