GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-11
Section 8-6
Electronic User Manual (EUM) Upgrade Procedure
8-6-1 Manpower
One Person, 15 min.
8-6-2 Tools
System DVD or Upgrade CD
8-6-3 Preparations
1.) Restart the system (turn off and on the system).
2.) Insert the System DVD or Upgrade CD into the drive.
3.) Press the UTILITIES
key on the control panel once to display the Utilities menu.
4.) Select the SYSTEM
item from the menu area to invoke the setup desktop on the screen.
5.) Select the
SERVICE page.
6.) Enter the password SHE, and click the ACCEPT
.
7.) The “Service Tools” menu appears on the screen.
NOTICE
The electronic user manual (EUM) must never be opened (by pressing the F1 key on the keyboard)
after last restart! Even if closed again before installing the new EUM.
Figure 8-15 Service Tools
EUM button