GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-15
Section 8-8
Replacement or Activation of Options
Following Options are available:
•Real Time 4D
•DICOM
• CrossBeam-CRI* (XBeam Compound Resolution Imaging)
•VOCAL II*
• RT_4D_Biopsy*
• VCI* (Volume Contrast Imaging)
• SRI* (Speckle Reduction Imaging)
•STIC*
• Tomographic Ultrasound Imaging* (T.U.I.)
NOTE: Additional option fields are not yet implemented in the Voluson® 730Pro / 730ProV.
8-8-1 Operation for activating Options
1.) Press the UTILITIES key on the control panel once to display the Utilities menu.
2.) Select the SYSTEM
item from the menu area to invoke the setup desktop on the screen.
3.) Select the
OPTION page where you can see which options are installed in the system.
• D = Demo (Option is activated for demo and expires on the date shown in the “Valid” column
• I = Inactive (Option is not activated)
• P = Permanent (Option is permanently activated [purchased] )
NOTICE
Not all Options are available on the Voluson® 730Pro V (marked with * in this manual).
BT Version:
Some Software Options may be considered standard, or are only available at systems that were
upgraded to BT´05 (SW 5.x.x). For details and description please refer to Section 5-2-5 on page 5-15.
Figure 8-21 Option page in the System Setup
Option page of
System type:
Voluson 730Pro
Option page of
System type:
Voluson 730Pro V