GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-19
8-9-2-4 Backup all Exams of the “old” system (cont’d)
7.) Click on the close button and then select the SETTINGS
icon on the left side of the screen.
8.) In the displayed window click on the CHANGE BACKUP FOLDER ON NETWORK DRIVE
button.
The first line in the dialog (see: Figure 8-23) displays the name (e.g., serial number A09008) of the
“Backup Folder” used for storing data on, and reading data from the network drive.
9.) Notice this name of the “old” system´s Backup Folder (e.g., A09008) and then click CANCEL
.
8-9-2-5 Restore all Exams (of the “old” system) to the “new” system
1.) On the system (= "new" system), that should receive data and images, press the SONOVIEW
key
on the Control panel.
2.) Select the SETTINGS
icon on the left side of the screen.
3.) In the displayed window click on the CHANGE BACKUP FOLDER ON NETWORK DRIVE
button.
4.) Choose folder of the “old” system (e.g., A09008) from the drop-down list and click on RENAME
.
Figure 8-23 Backup Folder on Network Drive
Figure 8-24 Step 1.) to 3.)
Figure 8-25 Select folder of “old” system and rename it
Name of the
Backup Folder on
Network Drive
Sonoview
key
Settings
icon
Change Backup Folder on Network Drive
Rename
button
choose folder of
“old” system
e.g., A09008