GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
8-20 Section 8-9 - Transfer of Patient Database and Images from System-to-System
8-9-2-5 Restore all Exams (of the “old” system) to the “new” system (cont’d)
5.) Close the “Settings” window with the OK
button.
6.) Click on the OPEN
icon on the upper left side of the screen and choose “Drive Network”.
7.) Using the
TRACKBALL, and the right trackball key SET, select the first exam of the list.
8.) To select all available exams, click on the SELECT TO END
button.
9.) Click on the BACKUP / RESTORE
button.
10.)In the “Backup / Restore” window click on the RESTORE
button.
NOTICE
The folder which is selected to be renamed (e.g., A09008, that contains the backup data from a different
system - in this case the backup data of the “old” system), it is not copied, but simply renamed.
Thus, the same data can be shared between two systems by renaming the respective backup folders
to the serial number of the accessing system (in this case the “new” system).
Figure 8-26 Step 5.) to 9.)
NOTICE
If an exam is about to be restored that already exists on the hard disk of the “new” system, a dialog
shows the Patient Name and Patient ID and asks for the action to be taken.
Figure 8-27 Identical exam already exists
Open
Backup / Restore
button
Select
to End
button
button
“Drive Network
all exams are selected
Close
button