GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
Console Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Peripheral Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Surface Probe Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Endocavity Probe
Leakage
Current Checks
Twice Annually
Measurement Accuracy
Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Probe/Phantom Checks
Check axial and lateral resolution (see Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks
of all probes Section 10-
5-2 on page 10-7
•
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-2 Periodic Maintenance Schedule
Item Service at Indicated Time D W M A Notes