GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
10-8 Section 10-5 - System Maintenance
10-5-2-2 Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
Refer to Table 3-8, “Approved Peripherals,” on page 43 for a list of approved peripherals.
10-5-3 Input Power
10-5-3-1 Mains Cable Inspection
10-5-4 Cleaning
10-5-4-1 General Cleaning
Table 10-6 Approved Peripheral/Hardware Option Functional Checks
Step Item Description
1VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
2 B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3 Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4
Line Printer /
Bluetooth Printer
Verify hardcopy output of the Line or Bluetooth printer. Clean heads and covers if necessary.
5 DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
6 Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
7
DVD and/or
MO-Drive
Verify that the DVD-drive and/or optional MO-drive reads/writes properly
(export/recall images in Sonoview).
8 Modem
Verify modem remote connection.
9ECG
Verify basic operation with customer.
Table 10-7 Mains Cable Inspection
Step Item Description
1 Unplug Cord
Disconnect the mains cable from the wall and system.
2 Inspect
Inspect it and its connectors for damage of any kind.
3 Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4 Verify
Inlet connector retainer is functional.
Table 10-8 General Cleaning
Step Item Description
1 Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Be careful not to get the cloth too wet so that moisture does not enter the console.
2 Probe Holder
Clean probe holders (they may need to be soaked to remove excess gel).
3 Monitor
Clean Top surface and Monitor and with a fluid detergent in warm water on a soft, damp cloth.