GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
10-10 Section 10-5 - System Maintenance
10-5-6 Optional Diagnostic Checks
Optionally you can access the diagnostic software as described in Chapters 5 or 7.
View the error logs and run desired diagnostics.
10-5-7 Probe Maintenance
10-5-7-1 Probe Related Checks
10-5-7-2 Basic Probe Care
The Basic User Manual provides a complete description of probe care, maintenance, cleaning and
disinfection. Ensure that you are completely familiar with the proper care of GE probes.
Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of
Voluson® 730Pro / 730ProV for more details. Failure to follow these precautions can result in serious
injury and equipment damage. Failure to properly handle or maintain a probe may also void its warranty.
Any evidence of wear indicates the probe cannot be used.
Do a visual check of the probe pins and system sockets before plugging in a probe.
10-5-7-3 Basic Probe Cleaning
Refer to the Basic User Manual of Voluson® 730Pro / 730ProV for details on cleaning.
NOTE: To help protect yourself from blood borne diseases, wear approved disposable gloves. These are made
of nitrile derived from vegetable starch to prevent allergic latex reactions.
NOTE: Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s
warranty. DO NOT soak or wipe the lens with any product not listed in the User Manual. Doing
so could result in irreparable damage to the probe. Follow care instructions that came with the
probe.
NOTE: Disinfect a defective probe before you return it. Be sure to tag the probe as being disinfected.
Table 10-10 Probe Related Checks
Step Item Description
1 Probe Holder
Clean probe holders (they may need to be soaked to remove excess gel).
2 Probes
Thoroughly check the system probe connectors and remove dust from inside the connector
sockets if necessary. Visually check for bent, damaged or missing pins.