3-2 Section 3-2 - Installation Reminders
3-2-2 Installation Warnings
1.) Since the Voluson® 730Pro / 730ProV weighs approximately 136 kg (300 lb.) without options, two
people are required to unpack it. Two people are required for installing any additional items in
excess of 16kg / 35 pounds (e.g., Monitor).
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
3.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to
acclimate before you turn it on. It requires one hour for each 2.5°C increment it's temperature is
below 10°C or above 40°C.
3-2-2-1 Brake Pedal Operation
3-2-2-2 Operator I/O Panel Position
If weight is placed on the Operator I/O Panel in it’s extended position the console could tip over.
CAUTION
Equipment damage possibility. Turning the system on without acclimation after arriving at site
may cause the system to be damaged.
Table 3-3 Acclimation Time
°C
60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40
°F
140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 -4 -13 -22 -31 -40
hrs
864200000002468101214161820
Figure 3-1 pulling, moving or lifting the system
WARNINGWARNING
REMEMBER: If the front wheels are engaged for transportation, pressing the release
brake pedals (brakes on front wheels under the foot rest) once disengages the lock.
WARNINGWARNING
The system should NOT be moved with the Operator I/O Panel extended.
Move the Operator I/O Panel to it’s centered and locked position.
WARNINGWARNING
Monitor mounting mechanism may break if not properly supported (e.g., with packing
foam) during transportation.
When pulling, moving or lifting the system,
grasp it only at the rear handle of the trolley and
the handle underneath the foot rest.
Do NOT pull or lift the system with the front handle
of the user interface (operator panel).
WARNING