GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
3-6 Section 3-4 - Preparing for Installation
Section 3-4
Preparing for Installation
3-4-1 Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order.
Report any items that are missing, back ordered or damaged.
3-4-2 Physical Inspection
3-4-2-1 System Voltage Settings
Verify that the scanner is set to the correct voltage. The Voltage settings for the Voluson® 730Pro /
730ProV Scanner is found on the identification plate, on the rear of the system.
3-4-3 EMI Protection
This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
Figure 3-4 Identification Plate
WARNINGWARNING
CONNECTING A Voluson® 730Pro / 730ProV SCANNER TO THE WRONG VOLTAGE
LEVEL WILL MOST LIKELY DESTROY THE SCANNER.