GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 3 - Installation 3-19
3-5-10-2 Disconnection (cont’d)
To stop the external device, select it and click the STOP
button.
A dialog shows which components will be stopped. To finish the process, click [OK].
Finally, a dialog shows that the device was stopped successfully.
The device can now be safely disconnected from the system.
By clicking OK
, the „Unplug or Eject Hardware“ dialog is active again.
Close this dialog by clicking CLOSE.
Afterwards select OK
to reboot the system.
Figure 3-20 Stop a Hardware device
Figure 3-21 Safe to Remove Hardware
WARNINGWARNING
Do not connect or disconnect any external USB-devices to or from the system while
scanning a patient! The appearing dialogs could distract you from the scan!
CAUTION
If the system’s DVD/CD writer or (optional) MO-drive was stopped by accident, simply stop the
external device as well and reboot the system.
During reboot, the DVD/CD writer and the MO-drive will be installed again.