GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
3-22 Section 3-6 - Completing the Installation
3-6-3 Transducer Connection
NOTE: Prior to connecting or disconnecting a probe, freeze the image.
It is not necessary to turn OFF power to connect or disconnect a transducer.
Connect a transducer to one of the three rightmost transducer receptacle as follows:
1.) Ensure that the transducer twist lock lever is at horizontal position.
2.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating
surface.
3.) Twist the transducer twist lock lever to vertical position to lock it in place. Twist the lever to the
horizontal position to disconnect the transducer.
4.) Open the right-hand side door, lay the cable into the intended cable holder and close the door.
Figure 3-25 Transducer Connection
CAUTION
If the cable spout on the right-hand door is missing, don’t pull the probe cable.
Please insert the spout in the designated place to avoid damage of the probe cable.