GE MEDICAL SYSTEMS - KRETZTECHNIK RAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
3-26 Section 3-7 - Printer Installation
3-7-2 Installing Digital Color Printer Sony UP-D21MD or UP-D23MD
1.) Power off/Shutdown the system as described in: Section 3-6-2 on page 3-21.
2.) Connect the printer as described on page 3-8 and reinstall the rear cover plate.
NOTE: For connection schemes refer to Section 3-5-8 on page 3-16 or see: Figure 3-5 on page 3-7.
NOTE: “Mouse functions” can be performed by using the trackball for moving the cursor.
“Normal select” (Click) = left/right trackball key ; “Opening a context menu” = upper trackball key
3.) Turn ON the printer, then switch ON the power of the system and wait till the system has booted.
NOTE: The power switch of the printer has to be in ON position before starting the system!
3-7-2-1 Install the UP-D21MD / UP-D23MD printer software/driver
Perform the following steps if this printer was never installed on the Voluson® 730Pro / 730ProV!
4.) Connect the USB cable to the printer and the system. The windows ‘Searching for Drivers …’,
‘Found new Hardware …’ and finally the following windows appear.
5.) Click the NEXT
buttons to start the Hardware Wizard and to locate the driver files.
NOTICE
Do not connect the USB-cable to the printer!
Figure 3-29 Found New Hardware Wizard
select “Search for a suitable driver for my
device (recommended)
Click on “Specify a location”

Table of Contents