GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 3 - Installation 3-27
3-7-2-1 Install the UP-D21MD / UP-D23MD printer software/driver (cont’d)
6.) Use the BROWSE
button to search the following path (see: Figure 3-27)
• C:\Utilities\PrinterDriver\SonyUP-D21MD or
• C:\Utilities\PrinterDriver\SonyUP-D23MD and then click OK
.
7.) Confirm the correct path and click NEXT
to install the driver. All necessary files are copied.
8.) The Message: Digital Signature Not Found appears. Click YES
.
9.) Confirm the installation by clicking FINISH
to close the Hardware Wizard.
10.)Close all open windows and restart the system (turn off and on the system).
Figure 3-30 Search for Network path
Figure 3-31 Confirm correct path and finish the Installation
NOTICE
Ver
After boot up of the system, verify the correct settings in the printer “Properties”, see: Section 3-7-4
"Adjustment of Printer Settings" on page 3-33.