GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 3 - Installation 3-41
3-8-3-1 How to enter Date and Time
Select the
GENERAL page in the System Setup see: Figure 3-49.
1.) Select the “Date Format” (only one can be active).
2.) Click the DATE/TIME
button to activate a sub dialog window to enter date, time and time zone.
3.) Click the TIME FORMAT
button to activate a sub dialog window to choose the preferred time
format.
4.) Click SAVE&EXIT
to save Settings and exit System Setup.
3-8-3-2 How to enter Hospital Name
Select the
GENERAL page in the System Setup see: Figure 3-49.
1.) Select the text box to enter a new “Clinic Name” with the keyboard.
2.) Click SAVE&EXIT
to save Settings and exit System Setup.
The clinic name will be copied into the Hospital ID in the information header.
3-8-3-3 How to change Language (and/or EUM Language)
Select the
GENERAL page in the System Setup see: Figure 3-49.
1.) Select the desired language from the pop-up menu.
2.) Click SAVE&EXIT
to save Settings and exit System Setup.
NOTE: After changing the language the system has to reboot.
Figure 3-49 System Setup - General page
BT Version:
Changing the EUM Language is only possible at systems that were upgraded to BT´05 (SW5.x.x).