GE MEDICAL SYSTEMS - KRETZTECHNIK DRAFT VOLUSON® 730PRO / 730PROV (BT´04)
D
IRECTION KTI105947, REVISION 2 SERVICE MANUAL
Chapter 3 - Installation 3-51
3-12-1 Map Network Drive
1.) Select the MAP NETWORK DRIVE button (in the System Setup - Network page, see: Figure 3-51)
to open a dialog where the system can be connected to a shared network drive of another server.
2.) Enter the name of the shared network folder in the „Network Folder Name“ field.
3.) Supply a valid user name and a password for this folder.
NOTE: If you check the „Automatic Reconnect“ box, the system tries to establish the connection
again when starting up.
Otherwise, the connection must be re-established manually after a shutdown or reboot.
4.) Select the CONNECT
button to establish the connection to the remote machine.
If successful, the DISCONNECT
button becomes active.
Figure 3-53 Map Network Drive window
NOTICE
If there is an error during the connection, a warning message appears inside the dialog.
In this case, please verify the data in the dialog.
NOTICE
If there already is a connection to the remote server, the CONNECT button is grayed.
To change the existing connection, first click on DISCONNECT
and then enter the new settings.
WARNINGWARNING
Please make sure that the server you are connecting to is trustworthy and reliable.
For details, contact your local system administrator.
If you backup Sonoview data to this server, all the patients’ demographic data will be
copied to this server!