8-20 Section 8-2 - System Software - Installation/Upgrade Procedure
8-2-2-5 System Software - Installation Procedure (FMI from external device) (cont’d)
19.)The system is rebooting into windows (Boot screen - Voluson).
20.)Please wait until all processes are finished. Finally the 2D screen is displayed on the monitor.
NOTE: If the BT warning dialog (Figure 8-27 below) appears, enter the Permanent key previously recorded
(refer to Figure 8-17: Option page (e.g., Voluson i) on page 8-15) and then click SUBMIT
.
21.)Remove the Stick from the USB port on the rear of the system.
22.)Reconnect the external devices, install all the printers and adjust the printer settings as described
in Section 3-7 "Printer Installation" on page 3-80.
23.)Check and match system settings in the
P1-P2-P3 page´s [Print], [Save] and [Export] tab.
24.)Confirm that the date and time are set correctly and that the Windows automatic DST feature is off.
Refer to Section Section 6-10 "Daylight Saving Time (DST) - New Dates" on page 6-12.
Figure 8-26 automatic processes are running
Figure 8-27 Enter Permanent key
enter “Permanent key”
then click [Submit]