IRECTION KTI106052, REVISION 10 SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
Probe/Phantom Checks
Check axial and lateral resolution (see Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen. NOTE!
The use of a Phantom is not required during
Preventive Maintenance. Customer may
use it as part of their Quality Assurance
Program tests.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks
of all probes Section 10-
5-2 on page 10-6
NOTE! The use of a Phantom is not required
during Preventive Maintenance. Customer
may use it as part of their Quality Assurance
Program tests.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-2 Customer Care Schedule
Item Service at Indicated Time D W M A Notes