DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
8-138 Section 8-42 - Replacement of the DRFM
6.) If “Operation was a success” is displayed on command window, patch is well installed.
7.) Shutdown to replace DRFM.
NOTE: After installing the software, please keep the software patch CD with the system for later usage.
For S10/S6/S8, please store the CD inside the rear cover. For P6/P8, please store the CD below the
top cabinet cover.
NOTE: In case of an inoperative system, please replace the DRFM-2 first before installing the software patch
CD. After installing the software patch CD, please re-start the system.
Figure 8-199 “Operation was a success” is displayed on command window