GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
10-34 Section 10-9 - When There's Too Much Leakage Current...
Type CF Patient Leakage Current Test - ECG Leads
ECG lead 1
ECG lead 2
ECG lead 3
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.01mA
Closed Normal Open 0.05mA
Closed Reversed Closed 0.01mA
Closed Reversed Open 0.05mA
Open Normal Closed 0.05mA
Open Reversed Closed 0.05mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.01mA
Closed Normal Open 0.05mA
Closed Reversed Closed 0.01mA
Closed Reversed Open 0.05mA
Open Normal Closed 0.05mA
Open Reversed Closed 0.05mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.01mA
Closed Normal Open 0.05mA
Closed Reversed Closed 0.01mA
Closed Reversed Open 0.05mA
Open Normal Closed 0.05mA
Open Reversed Closed 0.05mA